Child pages
  • Standardized Uptake Value (SUV)
Skip to end of metadata
Go to start of metadata

You are viewing an old version of this page. View the current version.

Compare with Current View Page History

Version 1 Next »

The standardized uptake value (SUV) formulas are used in positron emission tomography (PET) analysis. Radiotracer concentration can be measured for any area-based region of interest (ROI) such as circle, square, polygon, etc. The SUV formulas depend on certain attributes in the DICOM image without which SUV values cannot be calculated.

This application provides the following SUV types

  • SUV Body Weight (SUV bw)

  • SUV Lean Body Mass (SUV lbm)

  • SUV Body Surface Area (SUV bsa)

Note
N/A will be displayed in the event that SUV calculations are not available (e.g. due to missing DICOM attributes).

 

1. Required data in DICOM

  • General Tags
TagTagVRVMDescription
(0010,0040) CS1Patient’s Sex
(0010,1020) DS1Patient’s Size
(0010,1030) DS1Patient’s Weight
(0008,0021) DA1Series Date
(0008,0031) TM1Series Time
(0008,0022) DA1Acquisition Date
(0008,0032) TM1Acquisition Time
(0028,0051) CS1 - nCorrected Image
(0054,1102) CS1Decay Correction
(0054,1001) CS1Units
(0028,1052) DS1Rescale Intercept
(0028,1053) DS1Rescale Slope
  • Radiopharmaceutical Information Sequence
TagTagVRVMDescription
(0054,0016) SQ1Radiopharmaceutical Information Sequence
 (0018,1074)DS1Radionuclide Total Dose
 (0018,1078)DT1Radiopharmaceutical Start DateTime
 (0018,1072)TM1Radiopharmaceutical Start Time
 (0018,1075)DS1Radionuclide Half Life
  • Manufacture’s Tags
TagTagVRVMDescription
(0009,100D) TM1Scan Time for GE (private tag)
(7053,1000) DS1SUV Scale Factor for Philips
(7053,1009) DS1Activity Concentration Scale Factor for Philips

2. SUV formula

Main Formulas
  • Body Weight (SUV bw)
    SUV bw = [1]
  • Body Surface Area (SUV bsa)
    SUV bsa = [1]
  • Lean Body Mass (SUV lbm)
    SUV lbm = [2]
Factors of Formula
  • Body Surface Area
    BSA =
  • Lean Body Mass
    Male =
    Female =
  • Decay Correction
    DecayCorrection =
  • Time Gap
    TimeGap = Scan Time - Radiopharmaceutical Start Time

    If Decay Correction(0054,1102) is ADMIN, TimeGap is 0.
    If GE private tag has, Scan Time is with its value.

  • PET Image Pixels
    PETImagePixels =

[1]: Calculation of SUV bw and SUV bsa are according to the following equations, taken from Kim et al. Journal of Nuclear Medicine. Volume 35, No. 1, January 1994. pp164-167
[2]: Calculation of SUV lbm is according to Sugawara et al. Radiology, November 1999. pp 521-525

  • No labels